• GSI Novus GSI Novus Roll over image to zoom in
  • GSI Novus GSI Novus Roll over image to zoom in
  • GSI Novus GSI Novus Roll over image to zoom in
  • GSI Novus GSI Novus Roll over image to zoom in

GSI Novus

The new GSI Novus™ is a hand-held comprehensive newborn hearing screening instrument. The Novus features a touch-screen display and user-friendly software in a
compact hardware design. Novus may be configured with AABR, OAE or both which allows for seamless two stage infant screening.

• Combined AABR and OAE into one
• Touch screen display
• Customized risk factor management

• Compatible with tracking systems such as Hi-Track
• Wireless charging
• Data management software

General Specifications
Environmental conditions:
 + 15 °C… + 35 °C / + 59 °F… + 95 °F (operation)
 - 20 °C… + 50 °C / - 4 °F… + 122 °F (transport and storage)
 Maximum humidity 90 % (operation, non-condensating)
 Maximum humidity 95 % (storage, non-condensating)

Weight: 265 grams (with battery)

Dimensions: 158 mm x 83 mm x19mm
Display: 95 mm x 56 mm, color, 272 x 480 resolution
User Interface: Resistive touch screen
User Feedbac
k: Integrated speaker
Language Settings: English, default. (15 options)
Memory: 1GB
Data Interfaces: USB, Bluetooth®
Boot Up Time: <5 sec
Battery: Li-ion battery 44794; Capacity: 3.7V/3850 mAh
Warm Up Time: No warm-up time necessary after boot
Instrument Specifications – ABR
Test Signals: CE-Chirp®
Stimulus Rate: 88/sec left ear, 92.5/sec right ear
Stimulus Level: 35 dB nHL
Quality Control: Noise display
Instrument Specification – OAE
DPOAE
Stimulus Frequency Range: 1500 – 6000 Hz
Stimulus Level: 65 dB SPL
TEOAE
Stimulus Type: Non-Linear (according to IEC 60645-3)
Stimulus Frequency Range: 1000 – 4000 Hz
Stimulus Level: 83 dB peSPL, peak to peak calibrated, AGC controlled
Cradle
ELECTRICAL ISOLATION
DC Power In: 5V/1.6A
Power Supply: AC 100 – 240 V, ~ 50/60 Hz, 400mAA
Preamplifier
EEG Filter: 0.5 Hz – 5.0 kHz
Gain: 72 dB
CMRR: >110 dB at 100 Hz
Sample Rate: 22.05 kHz
Transducers
Radioear IP30 Insert Earphones
Probe for OAE and AABR testing
Printer (optional)
Type: Thermal
Connection: Bluetooth®
Battery: Lithium Ion, DC 7.4V, 1500 mAh
Charger: AC 100-250V, ~ 50/60 Hz, 1.0 A
Weight: 360 g / 12.7 oz
Paper: Thermal paper or labels